Event Text Entries

[182910869] No product has been returned for evaluation. Radiographs provided confirmed the alleged event. No root cause can be confirmed at this time.
Patient Sequence No: 1, Text Type: N, H10


[182910870] Information was received that a revision procedure was performed on (b)(6) 2020. As per the reporter, x-rays revealed a pin break. Reportedly, the patient had been struck across the back with a wooden pole. During the removal, bone had fused around the rod and the surgeon had to use a small mallet to encourage the rod to move and loosen. Once removed, it was noted that there was positioning /movement in the rod.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006179046-2020-00114
MDR Report Key9747852
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-24
Date of Report2020-02-24
Date of Event2020-01-07
Date Mfgr Received2020-01-27
Device Manufacturer Date2013-10-04
Date Added to Maude2020-02-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. TIARA RAE
Manufacturer Street101 ENTERPRISE SUITE 100
Manufacturer CityALISO VIEJO CA 92656
Manufacturer CountryUS
Manufacturer Postal92656
Manufacturer Phone6197314188
Manufacturer G1NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
Manufacturer Street101 ENTERPRISE SUITE 100
Manufacturer CityALISO VIEJO CA 92656
Manufacturer CountryUS
Manufacturer Postal Code92656
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Generic NameORTHOPEDIC INTERNAL FIXATION SYSTEMS, SPINAL
Product CodePGN
Date Received2020-02-24
Model NumberRA002-4545SL
Lot NumberA131004-02
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNUVASIVE SPECIALIZED ORTHOPEDICS, INC.
Manufacturer Address101 ENTERPRISE SUITE 100 ALISO VIEJO CA 92656 US 92656


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-24