ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM 24657

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-25 for ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM 24657 manufactured by Boston Scientific Corporation.

Event Text Entries

[180872944] It was reported that the device was difficult to remove. A 6x120x130 cm eluvia drug-eluting vascular stent system was selected for use in the left distal superficial femoral artery (sfa). The stent was successfully deployed. However, there was difficulty removing the stent delivery system from the patient. The tip of the catheter was stuck. The buddy wire technique was used and was able to remove the stent delivery system. The patient experienced no adverse effects and the procedure was completed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-01980
MDR Report Key9749994
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-25
Date of Report2020-02-25
Date of Event2020-01-31
Date Mfgr Received2020-02-03
Device Manufacturer Date2019-04-11
Date Added to Maude2020-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetBALLYBRIT BUSINESS PARK
Manufacturer CityGALWAY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Generic NameSTENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Product CodeNIU
Date Received2020-02-25
Model Number24657
Catalog Number24657
Lot Number0023728136
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.