COOK CHEST DRAIN VALVE N/A C-CDV-1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-25 for COOK CHEST DRAIN VALVE N/A C-CDV-1 manufactured by Cook Inc.

Event Text Entries

[183159651] (b)(6). Occupation: unknown. Pma/510(k) #: exempt. (b)(4). This report includes information known at this time. A follow-up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[183159652] It was reported that fluid from the patient was returning into a cook chest drain valve. As a result, the patient experienced "pulmonary expansion". The device was used both inside and outside of a healthcare facility during its implantation. During the patient's follow-up, it was noted that the valve was not waterproof. A presence of serious fluid and fibrin debris from the thorax was reported to have collected inside of the valve, which allowed the entry of air into the patient's thorax. This was validated by "immersion in water", which "went up in inspiration towards the pleural space". Additionally, there was a presence of clots obstructing the valve. No other adverse effects to the patient have been reported. Additional information regarding the patient and event has been requested but is currently unavailable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2020-00451
MDR Report Key9750600
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-25
Date of Report2020-02-25
Date Mfgr Received2020-02-06
Date Added to Maude2020-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. JENNIFER CANADA
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOOK CHEST DRAIN VALVE
Generic NameJOL CATHETER AND TIP, SUCTION
Product CodeJOL
Date Received2020-02-25
Model NumberN/A
Catalog NumberC-CDV-1
Lot NumberUNKNOWN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-25

© 2024 FDA.report