Event Text Entries

[180810208] Terumo has received the device for evaluation; however, the investigation has yet to be completed. Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[180810209] The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, the overpressure safety valve leaked. The valve was being used in an aortic root vent line and leaked during cardioplegia delivery. They were on regular integrate flow with the vent on and were told to turn off the vent. When the vent was turned it off, that's when the valve was noted to be leaking and it was like a steady stream coming out from the bottom of the valve. They took the 1st valve out and replaced it with a 2nd valve, which also leaked. The leaking stopped on the 3rd valve replacement. There was an approximate blood loss of 250 cc. There was a delay of 15 minutes. Product was changed out. Procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1124841-2020-00038
MDR Report Key9751655
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-25
Date of Report2020-03-12
Date of Event2020-01-31
Date Mfgr Received2020-02-26
Date Added to Maude2020-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CATHLEEN HARGREAVES
Manufacturer Street125 BLUE BALL ROAD
Manufacturer CityELKTON MD 21921
Manufacturer CountryUS
Manufacturer Postal21921
Manufacturer Phone8002837866
Manufacturer G1SAME
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVALVE, O.P.S BULK, N-S
Generic NameOVERPRESSURE SAFETY VALVE
Product CodeDTL
Date Received2020-02-25
Returned To Mfg2020-02-12
Model NumberLN130B
Catalog NumberN/A
Lot NumberXM07
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerTERUMO CARDIOVASCULAR SYSTEMS CORPORATION
Manufacturer Address125 BLUE BALL ROAD ELKTON MD 21921 US 21921


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-02-25