ZOLL IVTM THERMOGARD XP 8700-0650

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-25 for ZOLL IVTM THERMOGARD XP 8700-0650 manufactured by Zoll Circulation.

Event Text Entries

[184420372] The thermogard xp ivtm system (sn (b)(4)) involved in the reported complaint will not be returned for evaluation because the customer has resolved the problem by replacing the air tarp block. Therefore, service was not required to be performed by zoll.
Patient Sequence No: 1, Text Type: N, H10


[184420373] During set-up for patient use, the thermogard xp ivtm system ((b)(4)) displayed mid:09 (air trap assembly) error message upon powering on and the user observed burning smell from the ivtm system. In addition, the user noticed unknown liquid in the air trap block. Immediately, the ivtm system was replaced with another ivtm system and the patient treatment was completed without any interruptions. No consequences or impact to patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010617000-2020-00205
MDR Report Key9752467
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-25
Date of Report2020-02-25
Date of Event2020-02-03
Date Mfgr Received2020-02-03
Device Manufacturer Date2013-10-03
Date Added to Maude2020-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KIM NGUYEN
Manufacturer Street2000 RINGWOOD AVE,
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Manufacturer Phone4192922
Manufacturer G1ZOLL CIRCULATION
Manufacturer Street2000 RINGWOOD AVE.
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZOLL IVTM THERMOGARD XP
Generic NameTHERMAL REGULATING SYSTEM
Product CodeNCX
Date Received2020-02-25
Model Number8700-0650
Catalog Number8700-0650
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerZOLL CIRCULATION
Manufacturer Address2000 RINGWOOD AVE. SAN JOSE, CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.