OPTICROSS 8655

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-02-25 for OPTICROSS 8655 manufactured by Boston Scientific Corporation.

Event Text Entries

[180809737] Initial reporter city: (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[180809738] It was reported that catheter got stuck in the lesion and separation occurred upon removal. The 99% stenosed target lesion was located in a severly calcified below the knee artery by ipsilateral approach for endovascular treatment. An opticross imaging catheter was advanced for use. Following pre-dilation, the device was advanced, however, the catheter got stuck at the tip of the lesion. When trying to pull out, the catheter got separated at the blue shaft part about 28cm from the tip. Subsequently, it was retrieved by snaring. The device was completely removed from the patient. The procedure was completed with another of the same device. There were no patient complications reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-01898
MDR Report Key9752698
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-02-25
Date of Report2020-02-25
Date of Event2020-02-05
Date Mfgr Received2020-02-05
Device Manufacturer Date2019-11-26
Date Added to Maude2020-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street2546 CALLE PRIMERA
Manufacturer CityALAJUELA
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOPTICROSS
Generic NameCATHETER, ULTRASOUND, INTRAVASCULAR
Product CodeOBJ
Date Received2020-02-25
Model Number8655
Catalog Number8655
Lot Number0024837278
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.