SILKAM BLACK 3/0 (2) 75CM HR26 C0760412

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-26 for SILKAM BLACK 3/0 (2) 75CM HR26 C0760412 manufactured by B. Braun Surgical Sa.

Event Text Entries

[180919558] K990089 pma/510k: reported device not marketed in the u. S. , however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u. S. Analysis and results: there are no previous complaints of the same code-batch. We manufactured and distributed in the market 5,976 units of this code-batch. There are no units in our stock. We have received 60 closed samples for analysis. We have tested the knot pull tensile strength of all samples received and the results fulfil the requirements of the european pharmacopoeia (ep): 1. 20 kgf in average and 1. 07 kgf in minimum (ep requirements: 0. 92 kgf in average and 0. 31 kgf in minimum). Nevertheless, we have also tested the needle attachment strength of the samples received and the results do not fulfil the requirements of the european pharmacopoeia. Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfilled usp/ep and b. Braun surgical requirements. Knot pull tensile strength results before releasing the product were 1. 18 kgf in average and 1. 12 kgf in minimum and fulfilled ep requirements. Needle attachment strength results before releasing the product were 1. 06 kgf in average and 0. 94 kgf in minimum and fulfilled ep requirements. Final conclusion: taking into account that the results of samples received do not fulfil the specifications of the european pharmacopoeia/b. Braun surgical specifications concerning needle attachment strength, we conclude that the complaint is confirmed. Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product. Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions. Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
Patient Sequence No: 1, Text Type: N, H10


[180919559] It was reported that there was an issue with silkam black. The customer reported the needle is detached from the thread when loading the suture in the holder and during stomach surgery. The thread also broke when is loaded in the holder. It has happened to 3 different nurses and a surgeon.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003639970-2020-00123
MDR Report Key9755047
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-26
Date of Report2020-02-26
Date of Event2020-01-25
Date Mfgr Received2020-01-27
Date Added to Maude2020-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SILVIA SALUD ORUS
Manufacturer Street121 CARRETERA DE TERRASSA
Manufacturer CityRUBI, 08191
Manufacturer CountrySP
Manufacturer Postal08191
Manufacturer G1B. BRAUN SURGICAL SA
Manufacturer Street121 CARRETERA DE TERRASSA
Manufacturer CityRUBI, 08191
Manufacturer CountrySP
Manufacturer Postal Code08191
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSILKAM BLACK 3/0 (2) 75CM HR26
Generic NameOTHER SUTURE
Product CodeGAP
Date Received2020-02-26
Returned To Mfg2020-02-12
Model NumberC0760412
Catalog NumberC0760412
Lot Number619364
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerB. BRAUN SURGICAL SA
Manufacturer Address121 CARRETERA DE TERRASSA RUBI, 08191 SP 08191


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-26