STRAIGHT SUCTION 9733449

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-26 for STRAIGHT SUCTION 9733449 manufactured by Medtronic Navigation, Inc.

Event Text Entries

[184089718] Udi not available for this instrument. No parts have been received by the manufacturer for evaluation. Device manufacturing date is unavailable. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[184089719] Medtronic received information regarding a navigation system being used for a functional endoscopic sinus surgery (fess). It was reported that intra-operatively, the surgeon had difficulty verifying their straight suction. The surgeon needed to rotate the instrument after about ten attempts and finally could get it to verify. The procedure was completed using the navigation system and there was no reported impact to patient outcome. There was a reported delay to the procedure of less than one hour due to this issue. Medtronic received additional information that the cause or contributing factors related to the reported issue was that the instrument must have been slightly bent causing it to not verify easily.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1723170-2020-00651
MDR Report Key9757415
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-26
Date of Report2020-02-26
Date of Event2020-01-30
Date Mfgr Received2020-02-06
Date Added to Maude2020-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC NAVIGATION, INC
Manufacturer Street826 COAL CREEK CIRCLE
Manufacturer CityLOUISVILLE CO 80027
Manufacturer CountryUS
Manufacturer Postal Code80027
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTRAIGHT SUCTION
Generic NameINSTRUMENT, STEREOTAXIC
Product CodePGW
Date Received2020-02-26
Model Number9733449
Catalog Number9733449
Lot Number27167
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC
Manufacturer Address826 COAL CREEK CIRCLE LOUISVILLE CO 80027 US 80027


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.