KENDALL SCD 700 SERIES CONTROLLER, US M 29525B

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-02-26 for KENDALL SCD 700 SERIES CONTROLLER, US M 29525B manufactured by Covidien.

Event Text Entries

[181084488] The incident sample has been requested but to date has not been received for evaluation. If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted. As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
Patient Sequence No: 1, Text Type: N, H10


[181084489] The customer reported the device was attached to the end of the bed and when it was switched on, there was a bang and a flash where the cable connects to the unit. This resulted in the main circuit breaker within the local distribution board tripping. The customer also reported there was an unknown pinkish fluid dripping from the main cable connected to the machine. The unit had been in use in theater prior to this incident and had been returned to the maternity ward.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1282497-2020-08942
MDR Report Key9758796
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-02-26
Date of Report2020-02-26
Date of Event2019-10-31
Date Mfgr Received2020-02-20
Date Added to Maude2020-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJILL SARAIVA
Manufacturer Street15 HAMPSHIRE STREET
Manufacturer CityMANSFIELD MA 02048
Manufacturer CountryUS
Manufacturer Postal02048
Manufacturer Phone5086183640
Manufacturer G1COVIDIEN
Manufacturer Street15 HAMPSHIRE STREET
Manufacturer CityMANSFIELD MA 02048
Manufacturer CountryUS
Manufacturer Postal Code02048
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameKENDALL SCD 700 SERIES CONTROLLER, US M
Generic NameSLEEVE, LIMB, COMPRESSIBLE
Product CodeJOW
Date Received2020-02-26
Model Number29525B
Catalog Number29525B
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCOVIDIEN
Manufacturer Address15 HAMPSHIRE STREET MANSFIELD MA 02048 US 02048


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-26