CADD SOLIS HPCA PUMP 2110 21-2111-0300-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,user facility report with the FDA on 2020-02-26 for CADD SOLIS HPCA PUMP 2110 21-2111-0300-01 manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[181137012] One cadd solis hpca pump was received with no physical damage found on it. The event history log was reviewed and the reported problem was not found. Accuracy testing was conducted and the reported "failed accuracy" issue was unable to be confirmed. Delivery accuracy testing found the pump's average delivery error to be within the published specification of +/-6%. There was no fault found with the returned pump.
Patient Sequence No: 1, Text Type: N, H10


[181137013] Information was received indicating that a smiths medical cadd solis hpca pump's pm and volume reading are off. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-01471
MDR Report Key9759311
Report SourceCONSUMER,USER FACILITY
Date Received2020-02-26
Date of Report2020-02-26
Date of Event2020-01-14
Date Mfgr Received2020-01-27
Device Manufacturer Date2017-06-06
Date Added to Maude2020-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Manufacturer Phone3833310
Manufacturer G1SMITHS MEDICAL ASD, INC.
Manufacturer Street3350 GRANADA AVENUE NORTH SUITE 100
Manufacturer CityOAKDALE, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCADD SOLIS HPCA PUMP
Generic NamePUMP, INFUSION, PCA
Product CodeMEA
Date Received2020-02-26
Returned To Mfg2020-02-06
Model Number2110
Catalog Number21-2111-0300-01
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.