ONCONTROL COAXIAL BIOPSY TRAY 11/13 GA 9464-VC-006

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-02-27 for ONCONTROL COAXIAL BIOPSY TRAY 11/13 GA 9464-VC-006 manufactured by Teleflex Medical.

Event Text Entries

[181162750] (b)(4). The device has not been returned for investigation. Teleflex will continue to monitor and trend related events.
Patient Sequence No: 1, Text Type: N, H10


[181162751] It was reported that the patient was on the table and the needle was stuck in the bone and they could not get it out. The account ended up getting ortho to come out and get the needle out. Additional information from the physician indicates the patient suffered no adverse effects from the broken needle. We were able to have the orthopedic surgeon assist us during the procedure and bore out the bone around the broken oncontrol needle using a larger orthopedic drill. The needle came out easily. The bone lesion being biopsied was in the l iliac bone.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3011137372-2020-00074
MDR Report Key9763790
Report SourceHEALTH PROFESSIONAL
Date Received2020-02-27
Date of Report2020-02-19
Date of Event2020-02-19
Date Mfgr Received2020-02-19
Date Added to Maude2020-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEFFIE JEFFERSON
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194332672
Manufacturer G1TELEFLEX MEDICAL
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal Code27560
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameONCONTROL COAXIAL BIOPSY TRAY 11/13 GA
Generic NameBIOPSY NEEDLE KIT
Product CodeFCG
Date Received2020-02-27
Catalog Number9464-VC-006
Lot Number37581316
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressATHLONE


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-27