VDW SILVER RECIPROCATING MOTOR V041163000000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-02-27 for VDW SILVER RECIPROCATING MOTOR V041163000000 manufactured by Dentsply Vdw Gmbh.

Event Text Entries

[181635760] There has been a previous report received where stalling has caused file separation. Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur. As such, this event meets the criteria for reportability per 21 cfr part 803 please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply (b)(4). Battery (voltage breaks down under load) defective, replaced. Micromotor (bent shaft) defective, replaced. Sent in without foot control and power supply. Function test and test measurements without errors. Test carried out according to (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[181635761] In this event it was reported that a silver reciproc motor turns unbalanced and sometimes completely stops. No patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611053-2020-00041
MDR Report Key9764281
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-02-27
Date of Report2020-02-27
Date Mfgr Received2020-01-28
Date Added to Maude2020-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMR. KARL NITTINGER
Manufacturer Street221 W. PHILADELPHIA ST. SUITE 60W
Manufacturer CityYORK, PA
Manufacturer CountryUS
Manufacturer Phone8494424
Manufacturer G1ADVANCED TECHNOLOGY RESEARCH (A.T.R.) S.R.L.
Manufacturer StreetVIA SAN DONATO 1
Manufacturer CityPISTOIA, 51100
Manufacturer CountryIT
Manufacturer Postal Code51100
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVDW SILVER RECIPROCATING MOTOR
Generic NameCONTROLLER, FOOT, HANDPIECE AND CORD
Product CodeEBW
Date Received2020-02-27
Returned To Mfg2020-02-20
Model NumberNA
Catalog NumberV041163000000
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDENTSPLY VDW GMBH
Manufacturer AddressBAYERWALDSTRASSE 15 MUNICH, 81737 GM 81737


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-27

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.