MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional report with the FDA on 2020-02-27 for SECHRIST 3500 HL LOW-FLOW AIR OXYGEN BLENDER MIXER FOSC3000 manufactured by Sechrist Industries, Inc.
[181554661]
The complaint device was returned for evaluation. Initial visual inspection found no anomalies. Device evaluation identified that a knob was loose. The unit was re-calibrated. The alarm calibration test was ran and tested to oem specification. The root cause for the measurements being off was determined to be a loose know; however, it is unknown how the knob became loose. It was determined that this was not related to the previous repair. This type of event will continue to be monitored.
Patient Sequence No: 1, Text Type: N, H10
[181554662]
Reportedly, post repair, the device was reading off by 6%. There was no report of patient harm. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3007409280-2020-00041 |
MDR Report Key | 9765291 |
Report Source | CONSUMER,HEALTH PROFESSIONAL |
Date Received | 2020-02-27 |
Date of Report | 2020-02-04 |
Date of Event | 2020-01-30 |
Date Mfgr Received | 2020-02-04 |
Date Added to Maude | 2020-02-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | JENNA VARGAS |
Manufacturer Street | 212 AVENIDA FABRICANTE |
Manufacturer City | SAN CLEMENTE, CA |
Manufacturer Country | US |
Manufacturer Phone | 4495328128 |
Manufacturer G1 | AVANTE HEALTH SOLUTIONS |
Manufacturer Street | 212 AVENIDA FABRICANTE |
Manufacturer City | SAN CLEMENTE, CA |
Manufacturer Country | US |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | SECHRIST 3500 HL LOW-FLOW AIR OXYGEN BLENDER MIXER |
Generic Name | MIXER, BREATHING GASES, ANESTHESIA INHALATION |
Product Code | BZR |
Date Received | 2020-02-27 |
Returned To Mfg | 2020-02-14 |
Model Number | FOSC3000 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SECHRIST INDUSTRIES, INC |
Manufacturer Address | 4225 EAST LA PALMA ANAHEIM, CA US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-27 |