SECHRIST 3500 HL LOW-FLOW AIR OXYGEN BLENDER MIXER FOSC3000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional report with the FDA on 2020-02-27 for SECHRIST 3500 HL LOW-FLOW AIR OXYGEN BLENDER MIXER FOSC3000 manufactured by Sechrist Industries, Inc.

Event Text Entries

[181554661] The complaint device was returned for evaluation. Initial visual inspection found no anomalies. Device evaluation identified that a knob was loose. The unit was re-calibrated. The alarm calibration test was ran and tested to oem specification. The root cause for the measurements being off was determined to be a loose know; however, it is unknown how the knob became loose. It was determined that this was not related to the previous repair. This type of event will continue to be monitored.
Patient Sequence No: 1, Text Type: N, H10


[181554662] Reportedly, post repair, the device was reading off by 6%. There was no report of patient harm. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007409280-2020-00041
MDR Report Key9765291
Report SourceCONSUMER,HEALTH PROFESSIONAL
Date Received2020-02-27
Date of Report2020-02-04
Date of Event2020-01-30
Date Mfgr Received2020-02-04
Date Added to Maude2020-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJENNA VARGAS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE, CA
Manufacturer CountryUS
Manufacturer Phone4495328128
Manufacturer G1AVANTE HEALTH SOLUTIONS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSECHRIST 3500 HL LOW-FLOW AIR OXYGEN BLENDER MIXER
Generic NameMIXER, BREATHING GASES, ANESTHESIA INHALATION
Product CodeBZR
Date Received2020-02-27
Returned To Mfg2020-02-14
Model NumberFOSC3000
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSECHRIST INDUSTRIES, INC
Manufacturer Address4225 EAST LA PALMA ANAHEIM, CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-27

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.