ENSITE VELOCITY? DWS7 V.5.0.1 100149280 EE3000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-02-27 for ENSITE VELOCITY? DWS7 V.5.0.1 100149280 EE3000 manufactured by St. Jude Medical, Inc..

MAUDE Entry Details

Report Number2184149-2020-00029
MDR Report Key9765843
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2020-02-27
Date of Report2020-03-06
Date of Event2020-02-11
Date Mfgr Received2020-03-04
Device Manufacturer Date2018-04-04
Date Added to Maude2020-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL, INC.
Manufacturer StreetONE ST. JUDE MEDICAL DRIVE
Manufacturer CityST. PAUL MN 55117
Manufacturer CountryUS
Manufacturer Postal Code55117
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENSITE VELOCITY? DWS7 V.5.0.1
Generic NameCOMPUTER, DIAGNOSTIC, PROGRAMMABLE
Product CodeDQK
Date Received2020-02-27
Returned To Mfg2020-03-03
Model Number100149280
Catalog NumberEE3000
Lot Number6376989
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, INC.
Manufacturer AddressONE ST. JUDE MEDICAL DRIVE ST. PAUL MN 55117 US 55117


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-02-27

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.