MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-27 for ZOLL IVTM THERMOGARD XP 8700-0650 manufactured by Zoll Circulation.
[184419709]
Information received from the zoll service technician on 07 february 2020, the customer's biomed observed that the all four led's on the sensor block were lit. Following this, airtrap block was remove from the thermogard console (sn (b)(4)) and observed corrosion on the air trap block sensor's printed circuit board(pcb), thus confirming the reported complaint. Following this, the biomed ordered the airtrap block assembly to address the issue. Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for thermogard xp ivtm console with serial number (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[184419710]
During training, the air trap sensor assembly of the thermogard console (sn (b)(4)) was unable to recognize the air trap. The console did not display any alarm or error codes. No patient involvement.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3010617000-2020-00202 |
MDR Report Key | 9766232 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-02-27 |
Date of Report | 2020-02-27 |
Date of Event | 2020-02-06 |
Date Mfgr Received | 2020-02-06 |
Device Manufacturer Date | 2016-07-20 |
Date Added to Maude | 2020-02-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. KIM NGUYEN |
Manufacturer Street | 2000 RINGWOOD AVE, |
Manufacturer City | SAN JOSE, CA |
Manufacturer Country | US |
Manufacturer Phone | 4192922 |
Manufacturer G1 | ZOLL CIRCULATION |
Manufacturer Street | 2000 RINGWOOD AVE. |
Manufacturer City | SAN JOSE, CA |
Manufacturer Country | US |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ZOLL IVTM THERMOGARD XP |
Generic Name | THERMAL REGULATING SYSTEM |
Product Code | NCX |
Date Received | 2020-02-27 |
Model Number | 8700-0650 |
Catalog Number | 8700-0650 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ZOLL CIRCULATION |
Manufacturer Address | 2000 RINGWOOD AVE. SAN JOSE, CA US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-02-27 |