ICEFORCE 2.1 CX 90? CRYOABLATION NEEDLE FPRPR3602

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-27 for ICEFORCE 2.1 CX 90? CRYOABLATION NEEDLE FPRPR3602 manufactured by Galil Medical Ltd.

Event Text Entries

[182745282] The defective needle was not returned to the manufacturer, and the reported complaint could not be confirmed. The needle lot manufacturing records were reviewed, and no electrical deviations were found within the needle batch.
Patient Sequence No: 1, Text Type: N, H10


[182745283] Three iceforce needles were used in a cryoablation procedure. One needle caused twitching during the active thaw cycle. All the needles used during the procedure were turned on and off individually to identify the defective needle. The user switched to passive thaw with the defective needle, and the case was completed. The patient also incurred a skin burn with a separate needle during this procedure. Please see mdr # 9616793-2020-00040 for the report that describes this event. The defective needle was discarded, and therefore cannot be returned to the manufacturer for investigation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616793-2020-00039
MDR Report Key9766450
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-27
Date of Report2020-02-27
Date of Event2020-01-28
Date Mfgr Received2020-01-28
Device Manufacturer Date2019-10-02
Date Added to Maude2020-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SANDRA BAUSBACK-ABALLO
Manufacturer Street300 CONSHOHOCKEN STATE RD 300 FOUR FALLS CORPORATE CTR
Manufacturer CityWEST CONSHOHOCKEN, PA
Manufacturer CountryUS
Manufacturer Phone3311537
Manufacturer G1GALIL MEDICAL LTD.
Manufacturer Street1 TAVOR BUILDING PO BOX 224, INDUSTRIAL PARK
Manufacturer CityYOKNEAM, 2069203,
Manufacturer CountryIS
Manufacturer Postal Code2069203,
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameICEFORCE 2.1 CX 90? CRYOABLATION NEEDLE
Generic NameCRYOABLATION NEEDLE
Product CodeGEH
Date Received2020-02-27
Model NumberFPRPR3602
Catalog NumberFPRPR3602
Lot NumberT0207
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGALIL MEDICAL LTD
Manufacturer Address1 TAVOR BUILDING P.O. BOX 224, INDUSTRIAL PARK YOKNEAM, 2069203, IS 2069203,


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.