1/2 ML BD SAFETYGLIDE? INSULIN SYRINGE WITH ATTACHED NEEDLE 305934

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,foreign,other report with the FDA on 2020-02-28 for 1/2 ML BD SAFETYGLIDE? INSULIN SYRINGE WITH ATTACHED NEEDLE 305934 manufactured by Bd Medical - Diabetes Care.

Event Text Entries

[184892717] Date of event: unknown. The date received by manufacturer has been used for this field. Investigation summary: level a investigation - complaint evaluation / complaint history check for the event(s) that occurred. Severity: s_1__; occurrence: a complaint history check was performed and this is the 1st related complaint for scale marking light/illegible on lot # 8331575. Investigation summary: customer returned (1) 1/2cc, 8mm, 30g bd safetyglide insulin syringe in an open blister pack from lot # 8331575. Customer states that the plunger is broken. The returned syringe was examined and exhibited smeared ink printed on the barrel. Manufacturing (holdrege) will be notified of this issue. A review of the device history record was completed for batch# 8331575. All inspections were performed per the applicable operations qc specifications. There was one (1) notification [(b)(4)] noted for print out of registration. Investigation conclusion: based on the samples and/or photo(s) received the investigation concluded: confirmed: bd was able to duplicate or confirm the customer? S indicated failure (smeared scale markings). Complaints received for this device and reported condition will continue to be tracked and trended. Information will be captured on trend reports and monitored monthly. Our business team regularly reviews the collected data for identification of emerging trends. Root cause description: as per investigation completed by manufacturing under investigation child (b)(4), on 25feb2020, holdrege received a photo of a syringe 0. 5ml 30g rtw 8mm syringe from material 305934; batch 8331575. A visual evaluation of the syringe found double and shadow print throughout the numbers printed on the barrel. Process summary: the barrel printer applies ink from a substrate to a molded barrel that has been treated with corona to get a good adhesion of the ink to the molded barrel. The barrels are then transitioned to the oven for the ink to cure. The cured product exits the oven chute for transfer to the next operation. Root cause: dispatch 86780? On 17 dec 2019 double print coming off ct printer. Corrective action: print pads were changed. Rationale: based on the investigation, no additional investigation and no capa is required at this time.
Patient Sequence No: 1, Text Type: N, H10


[184892718] It was reported that 1/2 ml bd safetyglide? Insulin syringe with attached needle had scale marking issues. This was discovered before use. The following information was provided by the initial reporter: scale unclear.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1920898-2020-00182
MDR Report Key9767819
Report SourceCONSUMER,FOREIGN,OTHER
Date Received2020-02-28
Date of Report2020-02-14
Date of Event2020-02-12
Date Mfgr Received2020-02-12
Device Manufacturer Date2018-11-27
Date Added to Maude2020-02-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BD MEDICAL - DIABETES CARE
Manufacturer Street1329 WEST HIGHWAY 6
Manufacturer CityHOLDREGE NE 68949
Manufacturer CountryUS
Manufacturer Postal Code68949
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name1/2 ML BD SAFETYGLIDE? INSULIN SYRINGE WITH ATTACHED NEEDLE
Generic NameSYRINGE, ANTISTICK
Product CodeMEG
Date Received2020-02-28
Returned To Mfg2020-02-25
Catalog Number305934
Lot Number8331575
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBD MEDICAL - DIABETES CARE
Manufacturer Address1329 WEST HIGHWAY 6 HOLDREGE NE 68949 US 68949


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-02-28

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.