3 LESION NT1100? PAIN MANAGEMENT RF GENERATOR RFG-NT-1100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-28 for 3 LESION NT1100? PAIN MANAGEMENT RF GENERATOR RFG-NT-1100 manufactured by Neurotherm, Inc.

Event Text Entries

[181617809] The results/method and conclusion codes along with investigation results will be provided in the final report.
Patient Sequence No: 1, Text Type: N, H10


[181617810] The generator screen went black and eventually rebooted to the correct screen. In addition, the lesion graphical display stopped proceeding after a few seconds but the lesion time continue and temperature continued to display, and the impedance would sometimes drop to zero. Rebooting the generator did not resolve the issue. The generator had the same undiagnosed issue for roughly a year. The dates of the procedures are unknown. Approximately six procedures were cancelled due to the issue. Additional information is not available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002953813-2020-00008
MDR Report Key9769003
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-28
Date of Report2020-03-11
Date Mfgr Received2020-03-04
Date Added to Maude2020-02-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1NEUROTHERM, INC
Manufacturer Street600 RESEARCH DRIVE STE 1
Manufacturer CityWILMINGTON MA 01887
Manufacturer CountryUS
Manufacturer Postal Code01887
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3 LESION NT1100? PAIN MANAGEMENT RF GENERATOR
Generic NameGENERATOR, LESION, RADIOFREQUENCY
Product CodeGXD
Date Received2020-02-28
Returned To Mfg2020-03-03
Model NumberRFG-NT-1100
Catalog NumberRFG-NT-1100
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNEUROTHERM, INC
Manufacturer Address600 RESEARCH DRIVE STE 1 WILMINGTON MA 01887 US 01887


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2020-02-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.