MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2007-05-25 for * manufactured by Tontarra Medizintechnik Gmbh.

Event Text Entries

[16503820] Conclusion: no definite conclusion for the exact cause can be drawn. Above observations, however, leads us to the assumption, that the jaw part broke off due to overstressing or any kind of impact, when for example the instrument is dropped to the floor, or incautious handling during the cleaning, and sterilization process. Since our production documents indicate compliance with all specs, we feel that no corrective action has to be taken on our part. This incident is classified as not being security relevant as per the guideline of the european union: and it is not subject to be reported to the european authorities, because no pt injury incurred.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680515-2007-00005
MDR Report Key976999
Report Source08
Date Received2007-05-25
Date Mfgr Received2007-05-16
Device Manufacturer Date2005-01-01
Date Added to Maude2008-01-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag0
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location0
Manufacturer StreetDAIMLER STRASSE 15
Manufacturer CityWURMLINGEN 78573
Manufacturer CountryGM
Manufacturer Postal78573
Manufacturer Phone461965760
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name*
Generic Name*
Product CodeKOA
Date Received2007-05-25
Model Number*
Catalog Number*
Lot Number*
ID Number*
Device Eval'ed by MfgrY
Implant FlagN
Device Sequence No1
Device Event Key945033
ManufacturerTONTARRA MEDIZINTECHNIK GMBH
Manufacturer AddressDAIMLERSTRASSE 15 WURMLINGEN GM


Patients

Patient NumberTreatmentOutcomeDate
10 2007-05-25