ACRYSOF IQ TORIC SINGLEPIECE IOL SN6AT4 SN6AT4.110

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-02-28 for ACRYSOF IQ TORIC SINGLEPIECE IOL SN6AT4 SN6AT4.110 manufactured by Alcon Laboratories Ireland Ltd..

Event Text Entries

[184673189] This product malfunction was originally reported under an alternative summary report (e2000003). The sample has been returned and investigation was updated, which prompts evaluation and conclusion coding to be updated. As the alternative summary report e2000003 has been revoked, the updated information (product investigation and coding) is being sent via this 3500a report. The product was returned for analysis and the reported damage was observed. Additional observations were as follows: the iol was returned positioned correctly in the iol case. Solution is dried on the iol. The optic s scratched/marked rejectable. We are unable to determine the root cause for the reported complaint "lens desterilized, defective". The returned iol shows evidence of possible handling by the customer due to the presence of solution dried on the iol. In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed optic damage would not meet our current release criteria. Based on these investigation findings, we are unable to verify if the iol contributed to the event. All product and batch history records are quality reviewed prior to product release. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[184673190] A customer reported an intraocular lens (iol) was desterilized, defective. Additional information was requested.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612169-2020-00074
MDR Report Key9771101
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-02-28
Date of Report2020-02-28
Date Mfgr Received2020-02-25
Device Manufacturer Date2017-07-24
Date Added to Maude2020-02-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JONATHAN SCHLECH
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175514979
Manufacturer G1ALCON LABORATORIES IRELAND LTD.
Manufacturer StreetCORK BUSINESS&TECHNOLOGY PARK MODEL FARM ROAD
Manufacturer CityCORK 00000
Manufacturer CountryEI
Manufacturer Postal Code00000
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACRYSOF IQ TORIC SINGLEPIECE IOL
Generic NameLENS, INTRAOCULAR, TORIC OPTICS
Product CodeMJP
Date Received2020-02-28
Returned To Mfg2020-01-14
Model NumberSN6AT4
Catalog NumberSN6AT4.110
Lot Number21197190
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALCON LABORATORIES IRELAND LTD.
Manufacturer AddressCORK BUSINESS&TECHNOLOGY PARK MODEL FARM ROAD CORK 00000 EI 00000


Patients

Patient NumberTreatmentOutcomeDate
10 2020-02-28

© 2026 FDA.report