ENTERRA 37800

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-02-28 for ENTERRA 37800 manufactured by Medtronic Puerto Rico Operations Co..

Event Text Entries

[187656520] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[187656521] Additional information received from a healthcare professional (hcp) via a manufacturer representative (rep). It was reported that the cause of the event was thought to be the weight loss of the patient. Surgery to place a new battery was completed to resolve the issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004209178-2020-04343
MDR Report Key9771373
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-02-28
Date of Report2020-04-01
Date of Event2020-02-24
Date Mfgr Received2020-03-20
Device Manufacturer Date2015-06-04
Date Added to Maude2020-02-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLISA WOODWARD CLARK
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635263920
Manufacturer G1MEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer StreetROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK
Manufacturer CityJUNCOS PR 00777
Manufacturer CountryUS
Manufacturer Postal Code00777
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENTERRA
Generic NameINTESTINAL STIMULATOR
Product CodeLNQ
Date Received2020-02-28
Model Number37800
Catalog Number37800
Device Expiration Date2016-11-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer AddressROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK JUNCOS PR 00777 US 00777


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-02-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.