ICESPHERE? 1.5 CX 90? CRYOABLATION NEEDLE FPRPR3573

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-02-29 for ICESPHERE? 1.5 CX 90? CRYOABLATION NEEDLE FPRPR3573 manufactured by Galil Medical Ltd.

Event Text Entries

[186748273] Two cryoablation needles were used to treat a sclerotic metastatic thoracic vertebral body using bipedicular access. After the procedure, a thermal lesion to the spinal cord was discovered. The patient is paraplegic as a result of the lesion. The user believes the thermal lesion to be related to the cryoablation procedure. There was no reported malfunction with the cryoablation needles, and the needles will not be returned to the manufacturer for investigation. An anonymised transverse image of the needle placement was sent by the user to the manufacturer to estimate the expected posterior limit of the iceball. This mdr is being submitted for the second needle used during this procedure. Please see mdr # 9616793-2020-00045 for the first needle used during this procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616793-2020-00046
MDR Report Key9772514
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-02-29
Date of Report2020-02-29
Date of Event2020-01-29
Date Mfgr Received2020-01-30
Device Manufacturer Date2019-09-04
Date Added to Maude2020-02-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SANDRA BAUSBACK-ABALLO
Manufacturer Street300 CONSHOHOCKEN STATE RD 300 FOUR FALLS CORPORATE CTR
Manufacturer CityWEST CONSHOHOCKEN, PA
Manufacturer CountryUS
Manufacturer Phone3311537
Manufacturer G1GALIL MEDICAL LTD.
Manufacturer Street1 TAVOR BUILDING PO BOX 224, INDUSTRIAL PARK
Manufacturer CityYOKNEAM, 2069203,
Manufacturer CountryIS
Manufacturer Postal Code2069203,
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameICESPHERE? 1.5 CX 90? CRYOABLATION NEEDLE
Generic NameCRYOABLATION NEEDLE
Product CodeGEH
Date Received2020-02-29
Model NumberFPRPR3573
Catalog NumberFPRPR3573
Lot NumberT0102
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGALIL MEDICAL LTD
Manufacturer Address1 TAVOR BUILDING P.O. BOX 224, INDUSTRIAL PARK YOKNEAM, 2069203, IS 2069203,


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening; 2. Required No Informationntervention 2020-02-29