MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-02 for S8 ENT 9735669 manufactured by Medtronic Navigation, Inc.
        [181774501]
Device evaluation has not been done at the time of filing this report. If information is provided in the future, a supplemental report will be issued.
 Patient Sequence No: 1, Text Type: N, H10
        [181774502]
Medtronic received information regarding a navigation system used during a functional endoscopic sinus surgery. It was reported that the system displayed the following message on the registration screen,? An internal error has occurred and the application must restart. (code: #64)? Just after the 3d model was edited. The system restarted and the same issue kept occurring until the existing patient scans were deleted and re-copied onto the system via usb. This occurred pre-operatively and there was no delay to the procedure. There was no impact to the patient.
 Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1723170-2020-00691 | 
| MDR Report Key | 9774214 | 
| Report Source | COMPANY REPRESENTATIVE,FOREIG | 
| Date Received | 2020-03-02 | 
| Date of Report | 2020-03-16 | 
| Date of Event | 2020-02-26 | 
| Date Mfgr Received | 2020-03-09 | 
| Device Manufacturer Date | 2019-12-26 | 
| Date Added to Maude | 2020-03-02 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 3 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 3 | 
| Health Professional | 3 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 3 | 
| Event Location | 3 | 
| Manufacturer Contact | STACY RUEMPING | 
| Manufacturer Street | 7000 CENTRAL AVENUE NE RCW215 | 
| Manufacturer City | MINNEAPOLIS MN 55432 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 55432 | 
| Manufacturer Phone | 7635260594 | 
| Manufacturer G1 | MEDTRONIC NAVIGATION, INC | 
| Manufacturer Street | 826 COAL CREEK CIRCLE | 
| Manufacturer City | LOUISVILLE CO 80027 | 
| Manufacturer Country | US | 
| Manufacturer Postal Code | 80027 | 
| Single Use | 3 | 
| Previous Use Code | 3 | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | S8 ENT | 
| Generic Name | INSTRUMENT, STEREOTAXIC | 
| Product Code | PGW | 
| Date Received | 2020-03-02 | 
| Model Number | 9735669 | 
| Catalog Number | 9735669 | 
| Operator | HEALTH PROFESSIONAL | 
| Device Availability | N | 
| Device Age | DA | 
| Device Eval'ed by Mfgr | N | 
| Device Sequence No | 1 | 
| Device Event Key | 0 | 
| Manufacturer | MEDTRONIC NAVIGATION, INC | 
| Manufacturer Address | 826 COAL CREEK CIRCLE LOUISVILLE CO 80027 US 80027 | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 2020-03-02 |