UNKNOWN KNEE FEMORAL ADAPTOR BOLT UNK KNEE FEMORAL ADAPTOR BOLT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-02 for UNKNOWN KNEE FEMORAL ADAPTOR BOLT UNK KNEE FEMORAL ADAPTOR BOLT manufactured by Depuy Orthopaedics Inc Us.

Event Text Entries

[181579181] (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[181579182] Tc3 bolt and adaptor broke in patient.. Outcome was a revision for tc3 femoral component. There was metal debris in the joint and the sleeve was well fixed. Surgeon removed the femoral component after 30 minutes of work.. The 5 degree adaptor was still engaged in the sleeve which we were able to break the taper with slap hammer and five grip using the bolt.... There have been multiple cases with the same problem which i have not been a part of. Surgeon was able to re-engage morse taper with an srom hinge implant.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1818910-2020-06491
MDR Report Key9775009
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-02
Date of Report2020-02-12
Date of Event2020-02-12
Date Mfgr Received2020-02-12
Date Added to Maude2020-03-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street700 ORTHOPAEDIC DRIVE
Manufacturer CityWARSAW IN 465810988
Manufacturer CountryUS
Manufacturer Postal465810988
Manufacturer Phone6107428552
Manufacturer G1DEPUY ORTHOPAEDICS, INC. 1818910
Manufacturer Street700 ORTHOPAEDIC DR.
Manufacturer CityWARSAW IN 465810988
Manufacturer CountryUS
Manufacturer Postal Code465810988
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNKNOWN KNEE FEMORAL ADAPTOR BOLT
Generic NameUNK KNEE FEMORAL ADAPTOR BOLT
Product CodeHTN
Date Received2020-03-02
Catalog NumberUNK KNEE FEMORAL ADAPTOR BOLT
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEPUY ORTHOPAEDICS INC US
Manufacturer Address700 ORTHOPAEDIC DRIVE WARSAW IN 465810988 US 465810988


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-02

© 2026 FDA.report