OLYMPIC WARM SCALE (SMART SCALE, MODEL 23/28) 56323

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2008-01-08 for OLYMPIC WARM SCALE (SMART SCALE, MODEL 23/28) 56323 manufactured by Olympic Medical, A Division Of Natus Medical.

Event Text Entries

[745458] There was heat damage to the tray of an olympic warm scale (olympic smart scale, model 23/28). There was no patient injury because no patient was in the scale at the time of tray damage. No specific date for the heat damage event was provided. The scale was received for repairs on june 8, 2007. At the time this incident was reviewed, the device damage was determined to be caused by improper servicing and insufficient maintenance. However, this case was reviewed again in light of mdr manufacturing report #3018859-2008-00001 to be reportable by mdr.
Patient Sequence No: 1, Text Type: D, B5


[7932155] Evaluation summary & corrective action: the warm scale was returned to olympic medical for repair. It was determined that the heater coil was stretched out due to improper service; the heater was removed and stretched during hospital service. The stretched portion of the heater coil was nearest to the over-temperature thermistor sensor assembly; this resulted in overheating and damage to the tray. Note that the device was beyond its useful life. It was shipped on 2/25/98 and had a product life of 9 years so its end of life was 2/25/07; the incident occurred in approximately may of 2007. More importantly, the problem could have been detected earlier with annual heater calibration checks as specified in the instruction manual. However, no heater calibration checks were performed during the lifetime of the device. As part of mdr manufacturer report #3018859-2008-00001, a notification is being sent to all warm scale 23/28 customers to check the heater coil for uneven spacing and/or sagging in their warm scales. If they have any scales with a bunched, stretched, uneven or sagging heater coil, they should return the scale to olympic medical for service. This customer notification will also include a reminder to perform annual heater calibration as provided in the instruction manual.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2008-00002
MDR Report Key977599
Report Source05,06
Date Received2008-01-08
Date of Report2008-01-08
Date Mfgr Received2007-12-11
Device Manufacturer Date1998-02-25
Date Added to Maude2008-09-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone8004260353
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC WARM SCALE (SMART SCALE, MODEL 23/28)
Generic NamePATIENT SCALE (FRW)
Product CodeFRW
Date Received2008-01-08
Returned To Mfg2007-06-08
Model Number23
Catalog Number56323
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key980540
ManufacturerOLYMPIC MEDICAL, A DIVISION OF NATUS MEDICAL
Manufacturer AddressSEATTLE WA US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-01-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.