OLYMPIC WARM SCALE (SMART SCALE, MODEL 23/28) 56328

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2008-01-08 for OLYMPIC WARM SCALE (SMART SCALE, MODEL 23/28) 56328 manufactured by Olympic Medical, A Division Of Natus Medical.

Event Text Entries

[746031] There was heat damage to the tray of an olympic warm scale (olympic smart scale, model 23/28). There was no patient injury because no patient was in the scale at the time of tray damage.
Patient Sequence No: 1, Text Type: D, B5


[8087722] Evaluation summary & corrective actions: the warm scale was returned to olympic medical for investigation. It was determined that the heater coil was not evenly spaced. As a result, a portion of the coil sagged and electrically shorted out against the heater enclosure. This generated a hot spot in the heater coil; the additional heat caused tray damage. As part of the evaluation, the sagging, shorting and overheating was repeated on the customer's device. Note that the device was just beyond its useful life. It was shipped on 11/30/98 and had a product life of 9 years, so its end of life was 11/30/07; the incident occurred in late 2007. A corrective and preventative action (capa) has been opened to investigate potential corrective actions. This will include a notification sent to all customers to check the heater coil for uneven spacing and/or sagging in their warm scales; if they have any scales with a bunched, stretched, uneven or sagging heater coil, they should return the scale to olympic medical for service. The capa will also address a modification to the instruction manual to include annual inspection of the heater coils in addition to inspection following service of the heater. Finally, the capa will investigate possible modifications to the device itself.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2008-00001
MDR Report Key977600
Report Source05,06
Date Received2008-01-08
Date of Report2008-01-08
Date of Event2007-12-11
Date Mfgr Received2007-12-11
Device Manufacturer Date1998-11-30
Date Added to Maude2008-09-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone8004260353
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC WARM SCALE (SMART SCALE, MODEL 23/28)
Generic NamePATIENT SCALE (FRW)
Product CodeFRW
Date Received2008-01-08
Returned To Mfg2007-12-21
Model Number28
Catalog Number56328
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key980525
ManufacturerOLYMPIC MEDICAL, A DIVISION OF NATUS MEDICAL
Manufacturer AddressSEATTLE WA US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-01-08

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.