VITROS CHEMISTRY PRODUCTS BENZ REAGENT 680 1989

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2008-01-11 for VITROS CHEMISTRY PRODUCTS BENZ REAGENT 680 1989 manufactured by Ortho-clinical Diagnostics, Inc..

Event Text Entries

[745927] The customer obtained false negative vitros benz results on a single patient sample during a method comparison between the vitros 5, 1 fs chemistry system and a competitive system. Biased results of the direction and magnitude observed could lead to inappropriate physician action. The biased results were not reported and there was no report of patient harm as a result of this event. The report corresponds to ortho clinical diagnostics inc.
Patient Sequence No: 1, Text Type: D, B5


[8037821] The investigation determined that the vitros 5,1 fs was operating as intended. The false negative vitros benz results were obtained during a method comparison with a competitive system. The samples used were admixtures of multiple patient samples. The definitive root cause is unknown, however, it is possible that pre-analytical sample mixup occurred during the admixture process, although this could not be proven.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319808-2008-00007
MDR Report Key977692
Report Source05
Date Received2008-01-11
Date of Report2007-12-13
Date of Event2007-12-03
Date Mfgr Received2007-12-13
Device Manufacturer Date2007-08-01
Date Added to Maude2008-11-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOE FALVO
Manufacturer Street100 INDIGO CREEK DR.
Manufacturer CityROCHESTER NY 146265101
Manufacturer CountryUS
Manufacturer Postal146265101
Manufacturer Phone5854535735
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS, INC.
Manufacturer Street6010 LEE RD.
Manufacturer CityROCHESTER NY 14652310
Manufacturer CountryUS
Manufacturer Postal Code14652 3104
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS BENZ REAGENT
Generic NameIN-VITRO DIAGNOSTIC
Product CodeJXM
Date Received2008-01-11
Model NumberNA
Catalog Number680 1989
Lot Number1523-02-7634
ID NumberNA
Device Expiration Date2008-07-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key980163
ManufacturerORTHO-CLINICAL DIAGNOSTICS, INC.
Manufacturer Address100 INDIGO CREEK DR. ROCHESTER NY 146265101 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-01-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.