SACHSE COLD KNIFE, STRAIGHT 27068K

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-02 for SACHSE COLD KNIFE, STRAIGHT 27068K manufactured by Karl Storz Se & Co. Kg.

Event Text Entries

[188079772] Per the factory in (b)(6), upon evaluation the blade was found to be broken. There are visible discolorations on the shaft and on the breakage surface. The discoloration indicates corrosion. The broken off part of the blade was not returned. The device was manufactured in august 1989. Therefore it is suspected that the device was in use for more than 30 years. The root cause most probably is wear and tear as the device was manufactured in august 1989. Corrosion can weaken material. Due to wear the blade can get dull which will require higher force initiation during usage. This can cause damage and breakage of the blade. No indication for a manufacturing or material related issue was found during the investigation.
Patient Sequence No: 1, Text Type: N, H10


[188079773] Per the factory in (b)(6), allegedly there was an event occurred in (b)(6), during a procedure the blade from the optical urethrotome snapped inside the patient. The surgeon was satisfied there was no harm to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610617-2020-00031
MDR Report Key9777121
Report SourceUSER FACILITY
Date Received2020-03-02
Date of Report2020-02-27
Date of Event2019-11-28
Date Mfgr Received2020-02-12
Device Manufacturer Date1989-08-01
Date Added to Maude2020-03-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. SUSIE CHEN
Manufacturer Street2151 E. GRAND AVENUE
Manufacturer CityEL SEGUNDO, CA
Manufacturer CountryUS
Manufacturer Phone2188201
Manufacturer G1KARL STORZ SE & CO.KG
Manufacturer StreetDR.-KARL-STORZ -STRASSE 34 78532
Manufacturer CityTUTTLINGEN, GM,
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSACHSE COLD KNIFE, STRAIGHT
Generic NameURETHROTOME STRAIGHT COLD KNIFE
Product CodeEZO
Date Received2020-03-02
Model Number27068K
Catalog Number27068K
Lot NumberHV
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKARL STORZ SE & CO. KG
Manufacturer AddressDR.-KARL-STORZ-STRASSE 34 78532 TUTTLINGEN, GM, US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.