[181777081] Staff have been cut by sharp edges on the shelves of warmers that have holes punched out for circulation. During regular stocking or removing of items they can cut staff. Staff was cut on a model 5624. ~3 weeks later, staff was cut on a different 5624 unit. Hospital has large inventory of warmers. 4 were checked in icn where incidents happened and all were found to have sharp edges around the holes on the underside of all three shelves in each warmer. Manufacturer response for warmer, (brand not provided) (per site reporter). Getinge plans to come to the hospital next week to look at the shelves.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9779376
MDR Report Key9779376
Date Received2020-03-03
Date of Report2020-02-20
Date of Event2020-01-04
Report Date2020-02-21
Date Reported to FDA2020-02-21
Date Reported to Mfgr2020-03-03
Date Added to Maude2020-03-03
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMODELS 5524, 5618, 5624 WARMING CABINET
Generic NameWARMER, IRRIGATION SOLUTION
Product CodeLHC
Date Received2020-03-03
Model Number5624
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age5 YR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMAQUET CARDIOVASCULAR
Manufacturer Address45 BARBOUR POND DR WAYNE NJ 07470 US 07470


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-03