INTELLANAV MIFI OPEN-IRRIGATED ABLATION CATHETER 87047

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-03 for INTELLANAV MIFI OPEN-IRRIGATED ABLATION CATHETER 87047 manufactured by Boston Scientific Corporation.

Event Text Entries

[181831223] It was reported that irrigation difficulties occurred. During an percutaneous myocardial ablation procedure to treat atrial fibrillation in the right inguinal vein, an intellanav mifi open-irrigated ablation catheter was selected for use. While the catheter was being flushed at 60 milliliter per minute for priming, it was noted that one of the irrigation holes was flowing slower than expected. It was also noted that it was flowing forward rather than backwards. It was not reported whether any error messages were displayed. The procedure was completed by replacing the catheter. No patient complications were reported and the patient's current condition is fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-02462
MDR Report Key9779443
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-03
Date of Report2020-04-02
Date of Event2020-02-05
Date Mfgr Received2020-03-06
Device Manufacturer Date2019-09-30
Date Added to Maude2020-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTIMOTHY DEGROOT
Manufacturer Street4100 HAMLINE AVENUE NORTH DC A330
Manufacturer CitySAINT PAUL MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone6515826168
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street302 PARKWAY, GLOBAL PARK
Manufacturer CityLA AURORA - HEREDIA
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINTELLANAV MIFI OPEN-IRRIGATED ABLATION CATHETER
Generic NameCATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Product CodeOAD
Date Received2020-03-03
Returned To Mfg2020-02-20
Model Number87047
Catalog Number87047
Lot Number0024513949
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.