COMMON BILE DUCT EXPLORATION SET C-CDES-100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-03 for COMMON BILE DUCT EXPLORATION SET C-CDES-100 manufactured by Cook Inc.

Event Text Entries

[182930946] Concomitant products: laparoscopic maryland grasper, olympus ureteroscope. Pma/510(k): preamendment. This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[182930947] It was reported that the roadrunner pc wire guide included in a common bile duct exploration set stripped and became damaged during use in a laparoscopic cholecystectomy with an intraoperative cholangiogram and common bile duct exploration. The wire could have become stripped upon introduction into the working channel of the ureteroscope or during manipulation with the laparoscopic grasper used in the procedure. The damage occurred upon the 4th exchange. The 1st exchange was advancing the balloon dilation catheter over the wire, the 2nd exchange was the choloedochscope over the wire, the 3rd exchange was advancing the balloon dilation catheter of the wire, and the 4th exchange was the ureteroscope over the wire. The surgeon confirmed the wire guide stripped but no pieces came off or were left inside the patient. No other adverse effects were reported for this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2020-00529
MDR Report Key9779787
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-03
Date of Report2020-03-03
Date of Event2020-02-18
Date Mfgr Received2020-02-18
Device Manufacturer Date2020-01-28
Date Added to Maude2020-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. JENNIFER CANADA
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOMMON BILE DUCT EXPLORATION SET
Generic NameLQR DISLODGER, STONE BILIARY
Product CodeLQR
Date Received2020-03-03
Returned To Mfg2020-03-02
Model NumberNA
Catalog NumberC-CDES-100
Lot Number10082585
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-03