ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM 24653

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-03 for ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM 24653 manufactured by Boston Scientific Corporation.

Event Text Entries

[181837502] It was reported that in-stent restenosis occurred. The subject was enrolled in the (b)(6) study on (b)(6) 2017 and the index procedure was performed on the same day. The 95% stenosed target lesion was located in right distal superficial femoral artery (sfa). The target lesion was 140 mm long with a proximal reference vessel diameter of 6 mm and distal reference vessel diameter of 6 mm and was classified as tasc ii b lesion. Predilation was performed and a 6 mm x 150 mm study stent was implanted. Following post dilation, residual stenosis was 5%. On (b)(6) 2017, the subject was discharged. On (b)(6) 2019, the subject presented for the protocol specified 24 month follow-up visit. The subject was noted to have moderate stenosis in the stent on the right sfa. No action was taken to treat the event. At the time of reporting, the event was ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-02380
MDR Report Key9779795
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-03
Date of Report2020-03-03
Date of Event2019-10-30
Date Mfgr Received2020-02-12
Device Manufacturer Date2016-09-12
Date Added to Maude2020-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal Code55311
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Generic NameSTENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Product CodeNIU
Date Received2020-03-03
Model Number24653
Catalog Number24653
Lot Number0019834884
Device Expiration Date2018-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.