3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV) 3100A 768901-RFB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-03 for 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV) 3100A 768901-RFB manufactured by Vyaire Medical.

Event Text Entries

[184324302] Vyaire evaluate device onsite found out vent has no support arm. The circuit was disconnected where the support arm would be located. Excess water in the circuit cause map (main airway pressure) fluctuations. The customer was using low flow blender which is not recommended for ventilator use. Vyaire replaced circuit and retested system no other problems noted. The power module was replaced and performed 7 year pm. Completed performance test on the ventilator no other problems noted.
Patient Sequence No: 1, Text Type: N, H10


[184324303] The customer reported that the 3100a map (main airway pressure) was fluctuating and the amp reading is 2. The device was running on a patient and was transferred to another ventilator. Customer confirmed no patient harm in this reported event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2020-11481
MDR Report Key9780500
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-03
Date of Report2020-02-04
Date of Event2020-02-04
Date Mfgr Received2020-02-04
Device Manufacturer Date2013-11-26
Date Added to Maude2020-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MINDY FABER
Manufacturer Street26125 N. RIVERWOODS BLVD.
Manufacturer CityMETTAWA IL 60045
Manufacturer CountryUS
Manufacturer Postal60045
Manufacturer G1VYAIRE MEDICAL INC.
Manufacturer Street1100 BIRD CENTER DRIVE
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Generic NameVENTILATOR, HIGH FREQUENCY
Product CodeLSZ
Date Received2020-03-03
Model Number3100A
Catalog Number768901-RFB
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL
Manufacturer Address26125 N. RIVERWOODS BLVD. METTAWA IL 60045 US 60045


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-03