WM-NP2 WORKSTATION SET 4 (JP) K10021610

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-03-03 for WM-NP2 WORKSTATION SET 4 (JP) K10021610 manufactured by Keymed (medical And Industrial Equipment) Ltd..

Event Text Entries

[184889590] Olympus keymed have requested that the power cable is returned for further investigation. A follow up report will be submitted once an investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10


[184889591] The wm-np2 workstation was plugged into the facilities mains during preparation for use. Fire and smoke came from the plug on the power cable. One pin of the outlet terminal (plug) is missing - it appears to have melted. The operation was conducted using another workstation. There was no patient health damage.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611174-2020-00013
MDR Report Key9780641
Date Received2020-03-03
Date of Report2020-03-03
Date of Event2020-02-26
Date Mfgr Received2020-02-26
Date Added to Maude2020-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR PETER MORCOS
Manufacturer StreetKEYMED HOUSE, STOCK ROAD
Manufacturer CitySOUTHEND ON SEA, ESSEX SS25QH
Manufacturer CountryUK
Manufacturer PostalSS25QH
Manufacturer Phone0441702616
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWM-NP2 WORKSTATION SET 4 (JP)
Generic NameWM-*P2 SERIES OF ENDOSCOPY WORKSTATIONS
Product CodeFEM
Date Received2020-03-03
Model NumberK10021610
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerKEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
Manufacturer AddressKEYMED HOUSE, STOCK ROAD SOUTHEND-ON-SEA, ESSEX SS2 5QH UK SS2 5QH


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-03

© 2026 FDA.report