ACRYSOF IQ TORIC SINGLEPIECE IOL SN6AT3 SN6AT3.245

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other report with the FDA on 2020-03-03 for ACRYSOF IQ TORIC SINGLEPIECE IOL SN6AT3 SN6AT3.245 manufactured by Alcon Research, Llc - Huntington.

Event Text Entries

[184962855] Evaluation summary: the product was not returned for analysis. Product history and batch records were reviewed and documentation indicated the product met release criteria. The product investigation could not identify a root cause. There have been no other complaints reported in the lot number. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[184962856] A facility representative reported that during an intraocular lens (iol) implant procedure, after the lens was implanted, the doctor noticed a large scratch and cut out the lens and put in a new lens. The procedure was completed the same day. Additional information was received, that in the surgeon's opinion, the cause of the event, or what contributed to the event, may have been the lens loader. The lens loader was taken off of that tray to thoroughly clean. It was replaced with a new one. There are two medical device reports associated with this event. This report is associated with the intraocular lens (iol).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1119421-2020-00372
MDR Report Key9780821
Report SourceOTHER
Date Received2020-03-03
Date of Report2020-03-03
Date of Event2020-02-04
Date Mfgr Received2020-02-07
Device Manufacturer Date2019-06-14
Date Added to Maude2020-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JONATHAN SCHLECH
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175514979
Manufacturer G1ALCON RESEARCH, LLC - HUNTINGTON
Manufacturer Street6065 KYLE LANE
Manufacturer CityHUNTINGTON WV 25702
Manufacturer CountryUS
Manufacturer Postal Code25702
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACRYSOF IQ TORIC SINGLEPIECE IOL
Generic NameLENS, INTRAOCULAR, TORIC OPTICS
Product CodeMJP
Date Received2020-03-03
Model NumberSN6AT3
Catalog NumberSN6AT3.245
Lot Number12719518
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LLC - HUNTINGTON
Manufacturer Address6065 KYLE LANE HUNTINGTON WV 25702 US 25702


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-03