PERIPHERAL ROTALINK PLUS 29720

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-03 for PERIPHERAL ROTALINK PLUS 29720 manufactured by Boston Scientific Corporation.

Event Text Entries

[181830830] It was reported that blood leaked from the rotaburr sheath. The multiple tight stenosed target lesion was located in the moderately tortuous and approximately 250mm in length of extremely heavy calcified superficial femoral and popliteal artery. A 2. 00mm peripheral rotalink plus was selected for use. During procedure, after transversing about 100mm of the heavily calcified superficial femoral artery, it was noticed that the device stopped functioning correctly and blood was leaking out of the device. The rotalink plus was removed over the rotawire from inside the patient with no issues other than the device was leaking with blood/saline. This was after use of about 2-3 minutes in the patient. It was thought that the rotalink was punctured by a sharp calcification. The procedure was completed with another of the same device. No patient complications were reported and patient was good post procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-02430
MDR Report Key9780975
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-03
Date of Report2020-03-03
Date of Event2020-02-14
Date Mfgr Received2020-02-14
Device Manufacturer Date2019-09-24
Date Added to Maude2020-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORK LIMITED
Manufacturer StreetMODEL FARM ROAD
Manufacturer CityCORK
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePERIPHERAL ROTALINK PLUS
Generic NameCATHETER, PERIPHERAL, ATHERECTOMY
Product CodeMCW
Date Received2020-03-03
Model Number29720
Catalog Number29720
Lot Number0024484929
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.