Event Text Entries

[186565866] The device associated with this complaint was discarded by the customer. Since the device was not returned, an investigation could not be performed and a root cause could not be determined.
Patient Sequence No: 1, Text Type: N, H10


[186565867] During shift check, the autopulse platform displayed "replace battery" message upon insertion of the fully charged autopulse li-ion battery (sn: (b)(4)). The battery was placed into the autopulse multi-chemistry battery charger (mcc) for more than 24 hours; however, the status led on the charger showed amber light (test-cycle). There was no device malfunction reported on the autopulse platform, and the platform functioned as intended with different batteries. No patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010617000-2020-00209
MDR Report Key9781699
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-03
Date of Report2020-03-03
Date of Event2020-02-07
Date Mfgr Received2020-02-07
Device Manufacturer Date2017-03-29
Date Added to Maude2020-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KIM THOA NGUYEN
Manufacturer Street2000 RINGWOOD AVE,
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Manufacturer Phone4192922
Manufacturer G1ZOLL CIRCULATION
Manufacturer Street2000 RINGWOOD AVE.
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAUTOPULSE? LI-ION BATTERY
Generic NameLI-ION BATTERY, RECHARGEABLE
Product CodeMOQ
Date Received2020-03-03
Model NumberMODEL 100
Catalog Number8700-0752-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerZOLL CIRCULATION
Manufacturer Address2000 RINGWOOD AVE. SAN JOSE, CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-03