MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-03 for LEVEL 1 TRAUMA FAST FLOW SYSTEM H1200 8002950 manufactured by Smiths Medical Asd, Inc.
[181924411]
Evaluation results: one level 1 trauma fast flow system with a h-2 pressure cuff was received in fair condition. The device was manufactured in 2007. Visual inspection revealed that the h-31b air detector optical key sensor assembly was broken due to customer damage. The customer reported product problem was therefore confirmed. The h-31 optical key sensor assembly and mount block assembly were replaced as a result. A tempcheck test fixture was fitted on the device. The measured temperature was 41. 7 degrees which was within tolerance. The unit passed all functional and electrical tests.
Patient Sequence No: 1, Text Type: N, H10
[181924412]
It was reported that the air detector key sensor was broken. The product problem was observed during preventative maintenance. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3012307300-2020-01636 |
MDR Report Key | 9782476 |
Report Source | USER FACILITY |
Date Received | 2020-03-03 |
Date of Report | 2020-03-03 |
Date Mfgr Received | 2019-07-18 |
Device Manufacturer Date | 2007-05-31 |
Date Added to Maude | 2020-03-03 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | DAVE HALVERSON |
Manufacturer Street | 6000 NATHAN LANE NORTH |
Manufacturer City | MINNEAPOLIS, MN |
Manufacturer Country | US |
Manufacturer G1 | SMITHS MEDICAL ASD, INC |
Manufacturer Street | 3350 GRANADA AVENUE NORTH SUITE 100 |
Manufacturer City | OAKDALE, MN |
Manufacturer Country | US |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | LEVEL 1 TRAUMA FAST FLOW SYSTEM |
Generic Name | WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION |
Product Code | BSB |
Date Received | 2020-03-03 |
Returned To Mfg | 2019-07-30 |
Model Number | H1200 |
Catalog Number | 8002950 |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SMITHS MEDICAL ASD, INC |
Manufacturer Address | 6000 NATHAN LANE NORTH MINNEAPOLIS, MN US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-03 |