AMPLATZER PICCOLO OCCLUDER 9-PDAP-05-02-L

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,l report with the FDA on 2020-03-03 for AMPLATZER PICCOLO OCCLUDER 9-PDAP-05-02-L manufactured by Aga Medical Corporation.

Event Text Entries

[183946446] As reported in a research article, left pulmonary artery obstruction were noted during follow up after implantation with a 5-2 amplatzer piccolo and the device was explanted. The results of the investigation are inconclusive since the device was not returned for analysis. A review of the measurements as reported from the field was performed. Per the instructions for use, arten600042307 version a, "contraindications: weight <700 grams at time of the procedure" a piccolo device was not indicated for use in this patient. The cause of the reported event could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10


[183946447] Reference manufacturing report number: 2135147-2020-00090, 2135147-2020-00091, 2135147-2020-00092, 2135147-2020-00094. It was reported through a research article identifying amplatzer piccolo that may be related to a severe left pulmonary artery obstruction. Details are listed in the article, titled "device deformation and left pulmonary artery obstruction after transcatheter patent ductus arteriosus closure in preterm infants. " it was reported in the article that in (b)(6) 2016, 14 patients underwent transcatheter patent ductus arteriosus (pda) closure. A 21 days old patient that weigh 608 grams with pda diameter of 2. 1mm and pda length of 7. 7mm was implanted with a 5-2 amplatzer piccolo. During follow up, severe left pulmonary artery obstruction was observed and the device was removed from the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2135147-2020-00093
MDR Report Key9783105
Report SourceFOREIGN,HEALTH PROFESSIONAL,L
Date Received2020-03-03
Date of Report2020-03-03
Date Mfgr Received2020-03-02
Date Added to Maude2020-03-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAMELA YIP
Manufacturer Street5050 NATHAN LANE N
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1AGA MEDICAL CORPORATION
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal Code55442
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMPLATZER PICCOLO OCCLUDER
Generic NameTRANSCATHETER SEPTAL OCCLUDER
Product CodeMLV
Date Received2020-03-03
Model Number9-PDAP-05-02-L
Catalog Number9-PDAP-05-02-L
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAGA MEDICAL CORPORATION
Manufacturer Address5050 NATHAN LANE NORTH PLYMOUTH MN 55442 US 55442


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2020-03-03

© 2026 FDA.report