DRY ACID 132 GALLON UNIT, ASSY TO SHIP 160147 RTL160147

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-03-04 for DRY ACID 132 GALLON UNIT, ASSY TO SHIP 160147 RTL160147 manufactured by Concord Manufacturing.

Event Text Entries

[181908518] The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.
Patient Sequence No: 1, Text Type: N, H10


[181908519] A user facility biomedical technician (biomed) reported sparking coming from the fill valve on a dry acid mixer. Upon follow up, it was reported that a burning smell and a little bit of smoke also came from the machine. No further signs of burn damage were noted. The biomed replaced the fill valve and this did not fix the machine. The biomed then replaced the board in the display screen and the cable that connects to the fill valve, and that resolved the reported issue. After fixing the filling issue, an (unspecified) issue with the mixer motor surfaced. To resolve that issue, the biomed replaced the mixer motor. The biomed confirmed the machine was plugged into an appropriate outlet. At the time of follow up, the dry acid mixer was reported to be fully operational. There have been no further issues with the machine. The parts that were replaced have been discarded. There was no sample available to be returned for manufacturer evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2937457-2020-00426
MDR Report Key9784402
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-03-04
Date of Report2020-03-13
Date of Event2020-02-12
Date Mfgr Received2020-03-04
Device Manufacturer Date2016-12-09
Date Added to Maude2020-03-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMATTHEW AMARAL
Manufacturer Street920 WINTER ST
Manufacturer CityWALTHAM MA 02451
Manufacturer CountryUS
Manufacturer Postal02451
Manufacturer Phone7816999758
Manufacturer G1CONCORD MANUFACTURING
Manufacturer StreetDIRECTOR, QUALITY SYSTEMS 4040 NELSON AVENUE
Manufacturer CityCONCORD CA 94520
Manufacturer CountryUS
Manufacturer Postal Code94520
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDRY ACID 132 GALLON UNIT, ASSY TO SHIP
Generic NameDIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)
Product CodeKPO
Date Received2020-03-04
Model Number160147
Catalog NumberRTL160147
Device AvailabilityN
Device AgeMO
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCONCORD MANUFACTURING
Manufacturer AddressDIRECTOR, QUALITY SYSTEMS 4040 NELSON AVENUE CONCORD CA 94520 US 94520


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-04

© 2026 FDA.report