ARCHITECT STAT HIGH SENSITIVE TROPONIN-I 03P25-37

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-03-04 for ARCHITECT STAT HIGH SENSITIVE TROPONIN-I 03P25-37 manufactured by Abbott Ireland Diagnostics Division Longford.

Event Text Entries

[183385802] All available patient information was included. Additional patient details are not available. An evaluation is in process. A final report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[183385803] The customer observed a falsely depressed hstroponin result on the architect i2000sr analyzer. The following data was provided for a male patient diagnosed with acute myeloid leukemia: sid (b)(6) initial 138, repeats 20. 3, 84. 0, 74. 5, 175ng/l. The customer is using a cutoff of 26 ng/l. There was no negative impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005094123-2020-00064
MDR Report Key9785471
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-03-04
Date of Report2020-03-13
Date Mfgr Received2020-03-12
Device Manufacturer Date2019-10-26
Date Added to Maude2020-03-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTIAN LEE
Manufacturer Street100 ABBOTT PARK ROAD DEPT 09B9, BLDG CP01-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224668-294
Manufacturer G1ABBOTT IRELAND DIAGNOSTICS DIVISION LONGFORD
Manufacturer StreetLISNAMUCK CO. LONGFORD
Manufacturer CityLONGFORD NA
Manufacturer CountryEI
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT STAT HIGH SENSITIVE TROPONIN-I
Generic NameTROPONIN-I
Product CodeMMI
Date Received2020-03-04
Catalog Number03P25-37
Lot Number09200UI00
Device Expiration Date2020-04-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT IRELAND DIAGNOSTICS DIVISION LONGFORD
Manufacturer AddressLISNAMUCK CO. LONGFORD LONGFORD NA EI NA


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-04

© 2026 FDA.report