N
Patient 1
DEVICE RETURNED TO MANUFACTURER: THE COMET WAS RETURNED AND ANALYSIS WAS COMPLETED. THE TIP, DEVICE SHAFT, SENSOR PORT AND THE COEFFICIENT VALUES WERE EXAMINED FOR DAMAGE OR ANY IRREGULARITIES. THE WIRE SHOWED TWO KINKS LOCATED AT 176.5CM AND 180CM FROM THE TIP. THE TIP SHOWED BENDS. THERE WAS PEELED COATING AT THE 176.5CM LOCATION. THE OCC HANDLE WAS CONNECTED TO THE FFR LINK TO VERIFY THE SIGNAL STRENGTH. THE SIGNAL WAS PRESENT AS DESIGNED. THE SENSOR PORT SHOWED NO RESIDUE OF BODY FLUIDS. THE OCC HANDLE WAS THEN CONNECTED TO THE BENCH TOP TESTING EQUIPMENT AND THE WIRE WAS INSERTED INTO THE PRESSURE CHAMBER. THE PRESSURE WAS INCREASED TO VERIFY THE SENSOR WAS INDEED REACTING TO THE PRESSURE INCREASES AND DECREASES. THE PRESSURE SENSOR FUNCTIONED AS DESIGNED. THE COEFFICIENT VALUES WERE CONFIRMED TO BE PROGRAMMED. THE OCC HANDLE WAS AGAIN CONNECTED TO THE FFR LINK. THE DEVICE WAS THEN CONNECTED TO THE POLARIS (ILAB) TEST EQUIPMENT VIA BLUETOOTH SIGNAL. THE WIRE COMMUNICATED TO THE POLARIS SYSTEM AND ZEROED AS DESIGNED. THE WIRE WAS INSERTED INTO THE TEST PRESSURE CHAMBER AND THE WIRE TRANSFERRED A PD PRESSURE WAVEFORM TO THE POLARIS WHICH INDICATES A FUNCTIONING WIRE. THE WIRE WAS REMOVED FROM THE OCC HANDLE WITH NO ISSUES. INSPECTION OF THE REMAINDER OF THE DEVICE, APART FROM THE OBSERVED DAMAGE, REVEALED NO OTHER DAMAGE OR IRREGULARITIES.