MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign,health pr report with the FDA on 2020-03-04 for PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER 5MAXACE068 manufactured by Penumbra, Inc..
[188506626]
This device is available for return. A follow up mdr will be submitted upon completion of the device investigation.
Patient Sequence No: 1, Text Type: N, H10
[188506627]
The patient was undergoing a thrombectomy procedure in the right internal carotid artery (ica) using a penumbra system ace 68 reperfusion catheter (ace68). During the procedure, the physician steamed-shaped the tip of the ace68 and noticed that the ace68 seemed damaged; however, the physician continued to proceed with advancing the ace68 into the vessel. While advancing the ace68, the physician experienced resistance; and therefore, the physician decided to remove the ace68. Upon removal, it was noticed that the distal end of the ace68 was broken. The procedure was completed using a new ace68. There was no report of an adverse effect to the patient.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3005168196-2020-00293 |
MDR Report Key | 9786731 |
Report Source | DISTRIBUTOR,FOREIGN,HEALTH PR |
Date Received | 2020-03-04 |
Date of Report | 2020-02-10 |
Date of Event | 2020-01-29 |
Report Date | 2005-01-01 |
Date Reported to FDA | 2005-01-01 |
Date Reported to Mfgr | 2005-01-10 |
Date Mfgr Received | 2020-02-10 |
Date Added to Maude | 2020-03-04 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. VERONICA FARRIS |
Manufacturer Street | ONE PENUMBRA PLACE |
Manufacturer City | ALAMEDA CA 94502 |
Manufacturer Country | US |
Manufacturer Postal | 94502 |
Manufacturer Phone | 5107483200 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER |
Generic Name | NRY |
Product Code | NRY |
Date Received | 2020-03-04 |
Model Number | 5MAXACE068 |
Catalog Number | 5MAXACE068 |
Lot Number | F84790 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PENUMBRA, INC. |
Manufacturer Address | ONE PENUMBRA PLACE ALAMEDA CA 94502 US 94502 |
Product Code | --- |
Date Received | 2020-03-04 |
Device Sequence No | 101 |
Device Event Key | 0 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-04 |