16" CENTURY STERILIZER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-04 for 16" CENTURY STERILIZER manufactured by Steris Mexico, S. De R.l. De C.v..

Event Text Entries

[184922042] A steris service technician arrived onsite to inspect the 16" century sterilizer. While onsite, the user facility provided the technician with their cycle tapes. The technician reviewed the tapes and did not identify a cycle tape that aligned with the alarm subject of the reported event. The unit will alarm if the chamber does not reach the set temperature within the allotted time. The alarm indicates that the facility's steam supply is not providing sufficient steam to reach the set temperature. Additionally, the technicians spoke with the user facility's biomed department who stated that after reported event they had to "reset" the unit and returned the sterilizer to service. The technician performed an inspection and found no issue with the function or operation of the sterilizer. No repairs were required. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[184922043] The user facility reported via user facility medwatch report number (b)(4) that their 16" century sterilizer did not reach the proper temperature to continue the cycle.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005899764-2020-00009
MDR Report Key9787986
Report SourceUSER FACILITY
Date Received2020-03-04
Date of Report2020-03-04
Date Mfgr Received2020-02-04
Date Added to Maude2020-03-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DANIEL DAVY
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR, OH
Manufacturer CountryUS
Manufacturer Phone3927453
Manufacturer G1STERIS MEXICO, S. DE R.L. DE C.V.
Manufacturer StreetAVENIDA AVANTE 790 PARQUE INDUSTRIAL GUADALUPE
Manufacturer CityGUADALUPE, 67190
Manufacturer CountryMX
Manufacturer Postal Code67190
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name16" CENTURY STERILIZER
Generic NameSTERILIZER
Product CodeFLE
Date Received2020-03-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS MEXICO, S. DE R.L. DE C.V.
Manufacturer AddressAVENIDA AVANTE 790 PARQUE INDUSTRIAL GUADALUPE GUADALUPE, 67190 MX 67190


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-04

© 2024 FDA.report