MRI L/P PORT W/BROV 6.6F 0603880C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2020-03-04 for MRI L/P PORT W/BROV 6.6F 0603880C manufactured by Bard Access Systems.

Event Text Entries

[182154819] Photos were provided for review. The device has not been returned to the manufacturer for evaluation. As the lot number for the device was provided, a review of the device history records is currently being performed. The investigation of the reported event is currently underway. (expiration date: 08/2021).
Patient Sequence No: 1, Text Type: N, H10


[182154820] It was reported that approximately seven months post port device implant, the health care provider allegedly noticed subcutaneous swelling near the port site when injecting saline. It was further reported, that chest radiograph and angiogram demonstrated that the port catheter has allegedly fallen off the port body and migrated to the right atrium. Reportedly, the port body and detached catheter were both removed by interventional procedure. The patient is currently in good condition.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006260740-2020-00726
MDR Report Key9788567
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2020-03-04
Date of Report2020-03-04
Date of Event2019-12-19
Date Mfgr Received2020-02-09
Date Added to Maude2020-03-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJUDITH LUDWIG
Manufacturer Street1415 W. 3RD STREET
Manufacturer CityTEMPE AZ 85281
Manufacturer CountryUS
Manufacturer Postal85281
Manufacturer Phone4803032689
Manufacturer G1BARD REYNOSA S.A. DE C.V.
Manufacturer StreetBLVD MONTEBELLO #1 PARQUE INDUSTRIAL COLONIAL
Manufacturer CityREYNOSA TAMAULIPAS 88780
Manufacturer CountryMX
Manufacturer Postal Code88780
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMRI L/P PORT W/BROV 6.6F
Generic NameIMPLANTABLE PORT
Product CodeLJT
Date Received2020-03-04
Catalog Number0603880C
Lot NumberRECV2773
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBARD ACCESS SYSTEMS
Manufacturer Address605 N. 5600 W. SALT LAKE CITY UT 84116 US 84116


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2020-03-04