PACEART OPTIMA SYSTEM SOFTWARE POS12D18

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-04 for PACEART OPTIMA SYSTEM SOFTWARE POS12D18 manufactured by Medtronic, Inc..

Event Text Entries

[187224076] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[187224077] It was reported that the patient encounter association due to incoming data dialog resulted in the wrong patient data being assigned to an open patient record in the patient management database application. The clinic's information technology (it) support restored a copy of the production database to a new database, pointed to the test application server, opened the patient record, highlighted the patient encounter needed, select the save to disk file, obtained the pdd file and imported it into the production application, and edited the encounter, which corrected the issue. The application remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182208-2020-00434
MDR Report Key9788667
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-04
Date of Report2020-03-04
Date of Event2020-02-05
Date Mfgr Received2020-02-06
Date Added to Maude2020-03-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC, INC.
Manufacturer Street8200 CORAL SEA STREET NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal Code55112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePACEART OPTIMA SYSTEM SOFTWARE
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION, INDIRECT
Product CodeKRE
Date Received2020-03-04
Model NumberPOS12D18
Catalog NumberPOS12D18
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC, INC.
Manufacturer Address8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-04