CMAX SURGICAL TABLE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-04 for CMAX SURGICAL TABLE manufactured by Steris Corporation - Montgomery.

Event Text Entries

[184985930] A steris service technician arrived onsite following the reported event to inspect the cmax surgical table and found that wire from the battery fuse assembly was damaged. The user facility has opted to not repair the surgical table at this time and the table currently remains out of service. The cmax surgical table was placed into service in 2009 making it approximately 11 years old and is serviced and maintained by the user facility. The user facility did not disclose when the last preventive maintenance was completed for the table. The steris service technician counseled the user facility on the importance of regular preventive maintenance on their steris products. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[184985931] The user facility reported that after smelling smoke from the hallway, an employee found that their cmax surgical table had caught fire. A patient was not present on the table during the time of the reported event. No report of injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1043572-2020-00016
MDR Report Key9788767
Report SourceUSER FACILITY
Date Received2020-03-04
Date of Report2020-03-04
Date of Event2020-02-07
Date Mfgr Received2020-02-07
Date Added to Maude2020-03-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DANIEL DAVY
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR, OH
Manufacturer CountryUS
Manufacturer Phone3927453
Manufacturer G1STERIS CORPORATION - MONTGOMERY
Manufacturer Street2720 GUNTER PARK DRIVE EAST
Manufacturer CityMONTGOMERY, AL
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCMAX SURGICAL TABLE
Generic NameSURGICAL TABLE
Product CodeGDC
Date Received2020-03-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - MONTGOMERY
Manufacturer Address2720 GUNTER PARK DRIVE EAST MONTGOMERY, AL US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-04

© 2026 FDA.report