FUSION? ENT NAVIGATION SYSTEM 9733560XOM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-05 for FUSION? ENT NAVIGATION SYSTEM 9733560XOM manufactured by Medtronic Navigation, Inc.

Event Text Entries

[184086327] Other relevant device(s) are: product id: 9733467, serial/lot #: (b)(4). If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[184086328] Medtronic received information regarding a navigation system that was used during a functional endoscopic sinus surgery (fess) procedure. It was reported that the exam was not uploading properly. The 3d model was a white block and the exam only partially loaded. The exam included a small cross section of the nose. It was recommended that the site edit the scanner mask values which did not resolve the issues. It was also recommended that the site change the input settings in the overrides file. This resolved the white block 3d model issue however the exam was still not loading completely. Navigation was aborted due to this issue. There is no known impact on patient outcome.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1723170-2020-00757
MDR Report Key9792808
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-05
Date of Report2020-03-05
Date of Event2020-02-13
Date Mfgr Received2020-02-13
Date Added to Maude2020-03-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC NAVIGATION, INC
Manufacturer Street826 COAL CREEK CIRCLE
Manufacturer CityLOUISVILLE CO 80027
Manufacturer CountryUS
Manufacturer Postal Code80027
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFUSION? ENT NAVIGATION SYSTEM
Generic NameEAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Product CodePGW
Date Received2020-03-05
Model Number9733560XOM
Catalog Number9733560XOM
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC
Manufacturer Address826 COAL CREEK CIRCLE LOUISVILLE CO 80027 US 80027


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-05

© 2026 FDA.report