[182216027] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[182216028] Information was received from a healthcare professional (hcp) via a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator for gastric stimulation. It was reported that the hcp called the rep during a procedure to implant a new gastric stimulator. Prior to the procedure, the impedances were in the 700s [ohms]. After changing the stimulator, they were getting > 20k [ohms] for c/2, c/3, and 2/3. The rep indicated the hcp tried disconnecting the lead, wiping it down and then reconnecting it, but that didn? T resolve the issue. During the call with the rep, the hcp noted the impedances had come down so that c/2 was 6,614 [ohms], c/3 was 6,146 [ohms], and 2/3 was 672 [ohms]. The hcp reran impedances with the operating room (or) lights off and then c/2 was 693 [ohms], c/3 was 458 [ohms], and 2/3 was 647 [ohms]. The issue was noted to be resolved. No symptoms or further complications were reported or anticipated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004209178-2020-04698
MDR Report Key9793164
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-05
Date of Report2020-03-05
Date of Event2020-03-02
Date Mfgr Received2020-03-02
Device Manufacturer Date2015-06-04
Date Added to Maude2020-03-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLISA WOODWARD CLARK
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635263920
Manufacturer G1MEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer StreetROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK
Manufacturer CityJUNCOS PR 00777
Manufacturer CountryUS
Manufacturer Postal Code00777
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENTERRA
Generic NameINTESTINAL STIMULATOR
Product CodeLNQ
Date Received2020-03-05
Model Number37800
Catalog Number37800
Device Expiration Date2016-11-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer AddressROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK JUNCOS PR 00777 US 00777


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-05