INTELLANAV MIFI OPEN-IRRIGATED 87047

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-05 for INTELLANAV MIFI OPEN-IRRIGATED 87047 manufactured by Boston Scientific Corporation.

Event Text Entries

[182211599] Visual inspection revealed dried body fluid found on the handle, main shaft and distal end. The dried saline was found on the distal end and inside several irrigation ports. Electrical tests revealed that while manipulating the shaft, continuity checks revealed no electrical opens or shorts, as checked manually using a multi-meter and breakout box. All electrodes, sensor and thermocouple resistances measured in specification and were typical. Functional inspection revealed the steering knob and the tension control knob functioned properly on both lock and unlock positions. No abnormal resistance was felt when actuating the steering mechanism. The pressure leak test revealed the device's lumen was leak tested using an isaac pressure decay test system set to 107 psi, and a touhy bourst seal for sealing the irrigation holes and me's. The lumen pressure decay was measured three times, with re-seating of the tb seal between each test. Lumen pressure decay values were gross fail, gross fail and gross fail psi, which indicated a leak in the lumen. The x-ray inspection revealed that the x-ray found dried fluid debris adjacent to the cooling lumen inside the handle. Device dissection revealed the handle was opened to investigate a possible lumen leak. A crack was found in the adhesive that secures the irrigation tubing to the lumen. Dried saline was found inside the crack. Also found dried saline between several solder joints on the rear connector pcb. A metriq pump was connected to the luer fitting. Within seconds at a flow rate of 2ml/min, saline was observed exiting the crack in the adhesive.
Patient Sequence No: 1, Text Type: N, H10


[182211600] Reportable based on device analysis completed on 06feb2020. It was reported that there a high temperature error. During an denovo pulmonary vein isolation procedure to treat atrial fibrillation an nav mifi oi was selected for use. The nav mifi oi was used for about 15-20 minutes and rf was automatically stopped because of a high temperature error on the maestro. After clearing the error, the generator rebooted itself and got the "finding catheter" prompt. They power cycled the maestro and there was significant noise on the mifis. They unplugged and replugged the ablation cable and did not resolve the issue. They tried a new cable and there were no ablation signals coming from the catheter. They replaced the catheter which resolved the issues and the procedure was completed successfully. There was no patient complications. However, device analysis revealed a crack in the adhesive that joins the irrigation tubing to the cooling lumen allowed saline to leak into the interior of the handle.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-02617
MDR Report Key9793237
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-05
Date of Report2020-03-05
Date of Event2019-12-09
Date Mfgr Received2020-02-06
Device Manufacturer Date2019-05-21
Date Added to Maude2020-03-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTIMOTHY DEGROOT
Manufacturer Street4100 HAMLINE AVENUE NORTH DC A330
Manufacturer CitySAINT PAUL MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone6515826168
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street302 PARKWAY, GLOBAL PARK
Manufacturer CityLA AURORA - HEREDIA
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINTELLANAV MIFI OPEN-IRRIGATED
Generic NameCATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Product CodeOAD
Date Received2020-03-05
Returned To Mfg2020-01-07
Model Number87047
Catalog Number87047
Lot Number0023827401
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-05

© 2025 FDA.report
This site is not affiliated with or endorsed by the FDA.